– 10 –
10.0 ADJUSTING ADVANCED SETTINGS
There are several adjustable settings that can be changed from the factory defaults to
customize the WTT-110 to meet the exact requirements of the user. These settings are
saved in the microprocessor and are recalled each time the system is powered up.
10.1 Temperature Compensation (Default: SET=on)
In general, it is recommended to have the this function turned on as long as it
does not effect your measuring results. Only when measuring very minute
forces over a longer period of time at a slow rate and under stable environmental
conditions might it be useful to switch this off.
Switch off the instrument.
Press and hold PEAK- and MODE-buttons.
Press ON/OFF-button.
Wait until TRK oFF is displayed.
Release PEAK- and MODE-buttons.
10.2 Selecting the baud rate of the RS232C interface (Default: 2400)
On the WT-110RS model ONLY, the speed of the data transfer for the RS232C
interface can be adjusted as follows:
Switch off the instrument.
Press and hold RESET-button.
Press ON/OFF-button.
Wait until f01 is displayed, then release RESET-button.
Select function f04 by repeating to press the PEAK-button until f04 is shown
on the display.
Select appropriate baud rate (2400/4800/9600/19200) by repeating to press
the MODE-button.
Press RESET-button to save your settings.
10.3 Switching off Auto Power-Off function (Default: 10 min)
While operating under battery power the instrument will be automatically
switched off after approx. 10 minutes on inactivity. This Auto Power-Off func-
tion can be deactivated, which may become necessary when monitoring the
fluctuation of applied forces for a longer period of time. This function is
disabled when the instrument is powered by the AC-adaptor.
Switch off the instrument.
Press and hold RESET-button.
Press ON/OFF-button.
Wait until f01 is displayed, then release RESET-button.
Select function f03 by repeating to press the PEAK-button until f03 is shown
on the display.
Select setting (10/OFF) by repeating to press the MODE-button.
Press RESET-button to save your settings.
TRK
oFF
– 11 –
10.4 Change the refresh rate of the display (Default: 3/sec)
The force applied to the load cell of the instrument is internally processed at a
rate of 1000 Hz, which allows the precise capturing of the peak values. The read-
ings on the display, however, are refreshed 3 times/sec only for the convenience
of the human eyes. You may increase or decrease the factory settings for them to
match your personal demands.
Switch off the instrument.
Press and hold RESET-button.
Press ON/OFF-button.
Wait until f01 is displayed, then release RESET-button.
Select function f02 by repeating to press the PEAK-button until f02 is shown
on the display.
Select refresh-time (1/2/3/5/10/20 times/sec) by repeating to press the
MODE-button.
Press RESET-button to save your settings.