Series TT01 Digital Cap Torque Testers User’s Guide
19
3. To escape the Calibration menu at any time, press ESCAPE. The display appears as follows:
Selecting “Cancel” will revert back to the Calibration setup. Selecting “Exit w/o saving” will return
to the menu without saving changes.
4. After the number of calibration points has been entered, press ENTER. The display appears as
follows:
5. Place the tester on a level surface free from vibration, then press ZERO. The tester will calculate
internal offsets, and the display appears as follows:
If failed:
6. The following screen appears after the offsets have been calculated:
Attach weight fixtures (brackets, hooks, etc), as required. Do not yet attach any weights or apply
any calibration loads. Press ENTER.
CALIBRATION
NOT COMPLETE
Cancel
Exit w/o saving
CALIBRATION
CLOCKWISE
Attach necessary
weight fixtures,
then press ENTER.
CALIBRATION
OFFSET
Sensor failed
Analog failed
CALIBRATION
OFFSET
Sensor passed
Analog passed
CALIBRATION
OFFSET
Please wait…
CALIBRATION
OFFSET
Place torque sensor
horizontally, then
press ZERO.