PosiTest Pull-Off Adhesion Tester
INSTRUCTION MANUAL v. 3.0
Simple. Durable. Accurate.
PosiTest AT
(manual)
PosiTest AT-A
(automatic)
Vista de pagina 0
1 2 3 4 5 6 ... 10 11

Indice de contenidos

Pagina 1 - (automatic)

PosiTest Pull-Off Adhesion TesterINSTRUCTION MANUAL v. 3.0Simple. Durable. Accurate.PosiTest AT(manual)PosiTest AT-A(automatic)

Pagina 2 - Introduction

Page 11Technical DataTechnical DataPower Supply / Low Battery IndicatorPower Supply / Low Battery IndicatorPosiTest AT (manual)Power Source: AAA alkal

Pagina 3

Page 12DeFelsko’s sole warranty, remedy, and liability are the express limited warranty,remedy, and limited liability that are set forth on its websit

Pagina 4 - Step 3: Test Area Separation

IntroductionIntroductionThe portable, hand-operated PosiTest Pull-Off Adhesion Tester measures the forcerequired to pull a specified test diameter of

Pagina 5 - Step 4a: Pull-Off Test

Step 1: Dolly & Coating PreparationStep 1: Dolly & Coating PreparationDolly Preparation1. To remove oxidation and contaminants, place the incl

Pagina 6 - Troubleshooting (PosiTest AT)

Page 53. Gently push down on the dolly to squeeze out excess adhesive. Do not twist orslide the dolly back and forth on the coating as air bubbles may

Pagina 7 - Step 4b: Pull-Off Test

CLOSEPage 6Quick Guide(1) Open the pressure relief valve completely (turn counter clockwise)(2) Connect the actuator to the dolly (3) Close the pre

Pagina 8

Page 73. Close the pressure relief valve on the pump completely. (turn clockwise)As required, verify and adjust the dolly size by pressing thebutton.

Pagina 9

Page 8Dolly Size PSI Rates MPa Rates10 mm 125, 200, 400, 600, 725 1.00, 2.00, 3.00, 4.00, 5.0014 mm 60, 100, 200, 300, 360 0.40, 0.70, 1.40, 2.00, 2.5

Pagina 10 - Technical Data

Page 96. Remove the dolly from the actuator assembly.7. Readings may be stored into memory by pressing the button (pg. 10).All settings and

Pagina 11

Page 10Step 6: Storing Pull-Off values into MemoryStep 6: Storing Pull-Off values into MemoryThe PosiTest’s internal memory stores maximum pull pressu