Plug & Test
TM
Smart Force & Torque Sensors User’s Guide
14
11 SERIES R52 TORQUE SENSORS
11.1 Unpacking and Assembly
Carefully remove the sensor from the box. No assembly is required.
11.2 Overview
Designed for clockwise and counter-clockwise torque testing. The sensor
contains threaded holes on each side to permit mounting to a larger testing
system.
11.3 Specifications
Accuracy: ±0.35% of full scale
Safe overload: MR52-10Z – MR52-50Z: 300% of full scale
MR52-12 – MR52-100: 150% of full scale
Operating temperature: 40ºF – 100ºF [5ºC – 38ºC]
Operating humidity: 96% max. (non-condensating)
Weight: 1.2 lb [0.5 kg]
11.4 Dimensions
(in[mm])
2.87 [72.9]
2.00 [50.8]
0.57 [14.4]
45°
1/4" FEMALE SQUARE
#6-32 X4
#10-32 X4 REPEATED ON 2 SURFACES
Ø1.50 [38.1]
1.60 [40.6]
1.60 [40.6]
2.00 [50.8]
2.00 [50.8]
Ø1.76 [44.7]
24.0 [610] RETRACTED LENGTH
#10-32 X4
REPEATED ON 4 SURFACES
11.5 Capacity x Resolution
Model
No.
With Model 7i / 5i indicator With Model 3i indicator
ozFin lbFin lbFft gFcm kgFmm Nmm Ncm Nm ozFin lbFin kgFmm Ncm
MR52-10Z
10 x 0.005
-
-
700 x 0.5 7 x 0.005 70 x 0.05 7 x 0.005 - 10 x 0.01 -
7 x 0.005 7 x 0.005
MR52-20Z
20 x 0.01
-
-
1400 x 1
14 x 0.01 140 x 0.1 14 x 0.01 - 20 x 0.02 -
14 x 0.01 14 x 0.01
MR52-50Z
50 x 0.02
-
-
3600 x 2
36 x 0.02 350 x 0.2 35 x 0.02 - 50 x 0.05 -
36 x 0.05 35 x 0.05
MR52-12
-
12 x 0.005
1 x 0.0005
-
140 x 0.1 - 135 x 0.1 1.35 x 0.001 - 12 x 0.01
140 x 0.1 135 x 0.1
MR52-50
-
50 x 0.02
4 x 0.002
-
580 x 0.5 - 570 x 0.5 5.7 x 0.005 - 50 x 0.05
580 x 0.5 570 x 0.5
MR52-100
-
100 x 0.05
8 x 0.005
-
1150 x 0.5 - 1150 x 0.5 11.5 x 0.005 - 100 x 0.1
1150 x 1 1150 x 1
Comentarios a estos manuales