Checkline BL2 Manual de usuario Pagina 9

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 15
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 8
© Electromatic Equipment Co., Inc.
-20-
26.0 MOST COMMON REASONS FOR MISS READINGS WITH BL2 HAMMER
Product temperature out of application range
Material below 0°C resp. above +50°C (32 to 122 °F) may cause faulty
measurements.
Temperature difference between meter and sample
Please ensure that the device and the material under test are being stored at the
same temperature (+/-3°C) before measuring. A high temperature difference has
a negative effect on the stability of the measurement results.
Wrong calibration curve
Before you measure your sample, double check the correct selection of the
calibration curve.
Frozen wood or beetle infested wood
If you measure such products, the accuracy will decrease.
Measurement through bark
In fact of that the accuracy will decrease.
Direct solar radiation
27.0 MOST COMMON REASONS FOR MISS READINGS WITH FLW TESTER
Product temperature out of application range
Material below 0°C resp. above +40°C (32 to 104 °F) may cause faulty
measurements.
Temperature difference between meter and sample
Please ensure that the device and the material under test are being stored at the
same temperature (+/-3°C) before measuring. A high temperature difference has
a negative effect on the stability of the measurement results.
Water film at the measuring head
After measuring wet material a water film can arise on the sensor head. This
could lead to a too high result in the following measurement. After measuring
wet material clean both black plastic parts accurately with a dry cloth.
© Electromatic Equipment Co., Inc.
-9-
9.0 DESIGN FLW TESTER (ART. NO. 12119)
10.0 MEASURING PROCEDURE FLW TESTER
1. For a correct measurement please ensure that the
device has the same temperature than the material (+/-
3°C). For that reason, let your FLW tester adjust to the
surrounding temperature of the material before
measuring.
2. Switch on the device: Press the
key for three
seconds.
Measurin
g
ti
p
Handle
Sensor cable
Plu
g
Insertion
p
robe
Vista de pagina 8
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15

Comentarios a estos manuales

Sin comentarios