– 1 –TABLE OF CONTENTS1.0 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 022.0 Precautions . . . . . .
– 10 –4.1 Condition and Preparation of SurfacesIn any ultrasonic measurement scenario, the shape and roughness of the test surface are of paramount im
– 11 –5. Press the transducer against the probe-disc, making sure that thetransducer sits flat against the surface. The display should show some thick
– 12 –5.0 CALIBRATIONIn order for the TI-25DL-MMX to make accurate measurements, it must be set tothe correct sound-velocity for the material being me
– 13 –5.2 Calibration to a known velocityNote: This procedure requires that the operator know the sound-velocity ofthe material to be measured. A tabl
– 14 –4. Press the transducer against the sample piece, at the first/second calibration point, making sure that the transducer sits flat against the s
– 15 –6.0 QUICK START INSTRUCTIONS — STEEL THICKNESSThese Quick Start procedures are intended for those applications where the thick-ness of steel is
– 16 –7.0 CHANGING UNITS (IN/MM)The TI-25DL-MMX has the ability to display measurements in both the English(IN) and Metric (MM) systems. The following
TI-25DL-MMXULTRASONICTHICKNESS GAUGEOperating InstructionsCHECK•LINE®BY ELECTROMATICIncludes special versionModel TI-25DL-MMX-EXTELECTROMATIC E Q U I
– 2 –1.0 INTRODUCTIONThe TI-25M-MMX is a multi-mode Ultrasonic thickness gauge that is capable of measuring the thickness of various materials with ac
– 3 –3.0 OVERVIEW OF GAUGE3.1 GaugeProbeReceptaclesProbe Zero Test Plate and Battery Compartment CoverBacklit LCD DisplayMembraneKeypadProbe– 30 –18.0
– 4 –3.2 Contents Of KitThe TI-25M-MMX is supplied as a complete kit with the following:a. Gaugeb. Two (2) AA batteries (installed in gauge)c. Probe/
– 5 –3.4 The KeypadThe TI-25DL-MMX interacts with the operator through themembrane keypad and the LCD display. The functions ofthe various keys on the
– 6 –UP arrow is used to scroll through the various files, storage loca-tions, and functions of the data logger. Refer to page 22 for furtherinformati
These eight vertical bars form the StabilityIndicator. When the TI-25DL-MMX isidle, only the left-most bar and the under-line will be on. While the ga
– 8 –3.6 Probe Zero PlateWhen first connecting the probe supplied with the TI-25DL-MMX, the usershould perform a “Probe Zero” as described in Section
– 9 –4.0 MAKING MEASUREMENTSIn order for the transducer to do its job, there must be no air gaps between thewear-face and the surface of the material